植入材料浸提液中环氧乙烷迁移量 测试谱图:






Migration of ethylene oxide in the extract of implant material
Migration of ethylene oxide in the extract of implant material Details:
Name: Capillary Column
Specification: 30m*0.53mm*1.0um
Model: HH-FFAP
Application: Implantable medical devices, Polymer materials, Determination of toxic and hazardous substances in extracts. Part 10: Migration of ethylene oxide, Gas chromatography
Reagents and materials
Unless otherwise specified, the reagents used should be pure superior grade, and the test water should meet the requirements of seco
ndary water specified in GB/T 6682.
Ethylene oxide standard: ethylene oxide internatio
nal standard (50 000 mg/L) or ethylene oxide standard gas.
5 Reagent preparation
5.1 Preparation of ethylene oxide stock solution: if the internatio
nal standard ethylene oxide is used for preparation, mix the internatio
nal standard ethylene oxide upside down and take
Add 1.0 mL to a 100 mL volumetric flask pre-filled with ultrapure water, dilute to 100 mL, and store at 2℃~8℃. Concentration: 500 μg/mL.
5.2 If you use ethylene oxide standard gas to prepare the stock solution, take an externally dry 50 mL volumetric flask, add a
bout 30 mL water, and add a stopper.
Accurate weighing. Use a syringe to inject a
bout 0.6 mL of ethylene oxide, without the cork, shake it gently, change the cork, weigh, and weigh twice before and after
The difference is the weight of ethylene oxide co
ntained in the solution. Add water to the mark to make a solution co
ntaining a
bout 10 mg/mL of ethylene oxide, as a standard stock
liquid
Test spectrum of ethylene oxide migration in the extraction solution of implant material: