GDX-104色谱柱 测试谱图:

GDX-104 column
GDX-104 column details:
Non-methane hydrocarbons (NMHC) usually refer to all volatile hydrocarbons except methane (mainly C2~C8), also known as total non-methane hydrocarbons. The NMHC in the atmosphere exceeds a certain concentration, in addition to being directly harmful to human health, it can also produce photochemical smog under sunlight under certain conditions, causing harm to the enviro
nment and humans.
Tengzhou Haohan Chromatography Instrument Technology Service Co., Ltd. developed a GDX-104 stainless steel packed column, using "Air and Exhaust Gas Mo
nitoring and Analysis Method (Fourth Edition Supplement)" P587~P589 Total Hydrocarbon and Non-Methane Total Hydrocarbon Determination Method 1 ) The six-port valve quantitative loop sampling is used to determine the co
ncentration of non-methane total hydrocarbons in the ambient air and exhaust gas, and the instrument co
nditions are optimized.The detection limit of the method is 0.13 mg/m3, and the relative standard deviation of the standard sample total hydrocarbons is 0.4% The relative standard deviation of methane is 1.2%, the recovery rate of total hydrocarbons is 100.9~101.6%, and the recovery rate of methane is 102.4~102.6%. The method is simple in operation, fast in analysis and satisfactory in experimental results.
GDX-104 stainless steel packed column:
Details of GDX-104 are as follows:
name
composition
colour
Bulk density/g·ml-1
Specific surface
polarity
Zui high use temperature/℃
Main analytical uses
/m2·g-1
GDX-104
Copolymers such as divinylbenzene and styrene
translucent
0.22
590
Nonpolar
270
Gas analysis
Size: 3m*3mm
Particle size: 60-80 mesh
Application: Determination of total non-methane hydrocarbons
Application spectrum:
GDX-104 column test spectrum: